HOME | Products | Outgas Collector | FP-9000IIIDF

FP-9000ⅢDF for Flat panels

FP-9000ⅢDF
for the Flat Panel

 Outline
FP series is an outgas collector from flat panel samples such as LDC or PDP. FP-9000III is designed to collect surface outgas in the quartz oven with elevated temperature.
 
Feature
The oven plate has a space to place a flat panel sample and fix it firmly that makes accurate samplings possible.
 
Able to collect outgas from flat panel surface on the oven plate at maximum temp. at 400°C.
 
Able to collect outgas only from surface side, whilst outgas from backside is discharged completely by vacuum pump.
 
Oven temperature can go up to 400°C (when backing)
 
Chiller unit is the standard accessory for time saving of cool-down.
 
Able to install in a clean room

 
  
 

Safety System : Safety rock, Emergency stop, Leak detector, Overheat sensor stopper
Multi-step temperature, split sampling, sample tube cooler when sampling are available as option.
 
Touch panel is improved for user friendly operation.
 
 

 

Outline
FP series is an outgas collector from flat panel samples such as LDC or PDP. FP-9000III is designed to collect surface outgas in the quartz oven with elevated temperature.
 
Feature
The oven plate has a space to place a flat panel sample and fix it firmly that makes accurate samplings possible.
 
Able to collect outgas from flat panel surface on the oven plate at maximum temp. at 400°C.
 
Able to collect outgas only from surface side, whilst outgas from backside is discharged completely by vacuum pump.
 
Oven temperature can go up to 500°C (when backing)
 
Chiller unit is the standard accessory for time saving of cool-down.
 
Able to install in a clean room
 
Safety System : Safety rock, Emergency stop, Leak detector, Overheat sensor stopper
 
Multi-step temperature, split sampling, sample tube cooler when sampling are available as option.
 
Touch panel is improved for user friendly operation.
 
 

The oven plate to place flat panel samples
Place a flat panel sample (100mm x 70mm) on the oven plate center.

By vacuum pump, it enables to collect outgas only from surface side, whilst outgas from backside is discharged completely.
* The notch can be customized.

The oven plate to place flat panel samples
 
 
 
Place a flat panel sample (100 mm x 70 mm) on the oven plate center.  By vacuum pump, it enables to collect outgas only from surface side, whilst outgas from backside is discharged completely.  * The notch can be customized.

Specifications

Specifications

FP-9000ⅢDF
 Sampling method Hat shaped collecting method
 Sample size Smaller than 200 mm ID
 Oven material Stainless SUS316L
 Oven heating Surface and Bottom Individual heating method
 Heating temp. Max 400°C (up to 500°C for baking)
 Cooling Cooling water circulation
 Heater pipe Inside inert treated
 Purge gas N2 or He for general use, Flow rate up to 500 ml/min
 PAT cooling unit (option) Electronic cooling
 Overheat prevention sensor Power 15A Sensitivity current 30mA
 Emergency stop Available
 Safety system Safety interlock system
 Pressurization method Motor drive
 Dimensions 1140(W) x 1106(H) x 750(D) mm
 Power supply Single phase AC200V, 20A AC200V, 30A
 Weight 280 Kg

 
 

 

FP-9000ⅢDF
Sampling method Hat shaped collecting method
Sample size Smaller than 200 mm ID
Stainless SUS316L Stainless SUS316L
Oven heating Surface and Bottom Individual heating method
Heating temp. Max 400°C (up to 500°C for baking)
Cooling Cooling water circulation
Heater pipe Inside inert treated
Purge gas N2 or He for general use, Flow rate up to 500 mL/min
PAT cooling unit (option) Electronic cooling
Overheat prevention sensor Power 15A Sensitivity current 30mA
Emergency stop Available
Safety system Safety interlock system
Pressurization method Motor drive
Dimensions 1140(W) x 1106(H) x 750(D) mm
Power supply Single phase AC200V, 20A AC200V, 30A
Weight 280 Kg