Outline
Up to 40 samples in the Pyrofoils can be loaded onto the Pyrofoil
magazine. Major labor savings of routine analysis are provided.
Feature
• Forty (40) sample capacity.
• Oven temperature adjustable to 400°C.
• Routine analysis can be performed with labor saving
ease.
• Due to the pyrofoil system, samples, i.e. vulcanized
rubber, can be weighted accurately resulting in
highly reproducible quantitative analysis.
• Both solid and liquid samples can be analyzed
automatically.
• Any commercial GC can be interfaced.
Specification
♦Sampler Unit
Sample Capacity: 40 samples
Foil Induction: Pneumatic activated arm drops
the
foil into the RF chamber
Foil Ejection: Electromagnetic rod picks up the
spent foil and deposits it into the foil
trap
Dimension/weight:
254 W x 220 H x 400 D (mm)/3 Kg
♦GC Probe
Oven Temperature: Adjustable to 400°C, 300°C max
recommended for long run time.
Needle Temperature: Adjustable to 350°C, 300°C
max recommended for long run time.
♦Power Supply/ Controller
Pyrolysis: Curie point heating (Radio Frequency
Induction Heating)
RF Frequency: 600 KHz, 48 W
Power Supply: 600 KHz, 48 W
Pyrolysis Time: 1 to 9,999 seconds
Dimensions/ weight:
180 W X 320 H X 200 D (mm)/13.5 Kg
Power Requirements:
85 to 115 V, 8 A (Available for 220 V)
♦Flow Controller, FWC-33B
Pressure Regulator: Adjustable from 0.1 to 0.4
MPa
Purge Valve: Controlled by a 3 port solenoid valve
Dimensions/weight:
100 W X 150 H X 150 D (mm)/5 Kg
|



|