HOME > Products > Recycling Preparative HPLC > LC-9110II /9130II NEXT Recycling Preparative HPLC Japan Analytical Industry Co.,Ltd. Copy

LC-9110ⅡNEXT/9160ⅡNEXT

Improved High-end NEXT
NEXT for the next generation

Recommended configuration example
LC-9110II NEXT + JAIGEL-H Column (lD20mm x 2) + UV-Vis 4ch NEXT + RI-700NEXT
LC-9160II NEXT + JAIGEL-ODS Column + UV-Vis DAD NEXT + RI-700+ ClarityTM

Recycle with Non-diffusion (patent)
During extended recycling period, the recycled sample in the flow line may diffuse back into the solvent with the danger of contaminating the solvent reservoir. The N EXT series has non-diffusion device (patent) that prevents the sample from running back into the solvent line and damaging your sample,columns and detectors with back pressure.
Flash solvent function
Automatic flow line switching between the mobile phase and the column cleaning solvent is become reality.
Column cleaning after separation can be easily completed.

・Other features
Besides the recycle function is equipped, other various functions are equipped for your convenience, such as repeat injector, auto recycle, auto replay and fraction time auto adjusting.

・Large size LGD touch panel for easy operation
8.4 inch size touch panel screen enables easy and user-intuitive operation by simply touching icons at the fl ow channel diagram on the screen from start to collect.

Flow channel





・Auto cleanup function
Keeps the flow channels clean automatically. Solvent changeover when changing columns is also made much easily.

・Reliable solvent delivery peformance
Optimized serial double plunger pump dedicated for recycling is installed. Addition of digital flow rate compensation function enables reliable solvent delivery performance. The exact flow rate as setting value is guaranteed,regardless of column pressure.
250
・PC control function
With optional PC controling software Prepure-200 (or higher),it enables controlling LC-91N EXT remotely a nd collecting target peaks automatically with peak mode fraction.

・Injection overlap function
Enables to make second and subsequent injections without having to wait to complete previous operation.

Chromatogram data acquisition system

There are a few selection of software depends on your need and detector selection. UV-Vis 4ch NEXT works perfectly with JAI Scan and UV-Vis DAD NEXT works with ClarityTM. For other detectors,please choose from Prepure or JDS series.

Detectors

Two different types of detectors are both smal enough to fit into NEXT series body. They can be installed together and perform detection simultaneously.

UV Detector

As preparative dedicated detector, it has developed its stability and sensitivity can detect most organic compounds.

UV-254NEXT UV-370NEXT
UV-600NEXT
UV-VIS 4ch NEXT
Detection Fixed wavelength Variable wavelength Variable wavelength 4channel variable wavelenghth
Wave length 254nm 195~370nm
195~600nm
190~480nm
Measurement Method Single beam Double beam Single beam
 Flow Cell Cell Length: 0.3mm, Cell volume: 40µl
(Option: 1.0mm, 2.4mm)
Cell Length: 0.5mm,
Cell volume: 20µl
(Option: 0.2mm, 0.1mm)
Cell Length: 0.3mm, Cell volume: 40µl
(Option: 1.0mm, 2.4mm)

RI Detector

Stable baseline can be obtained quickly. Applicable for both preparative and analytical HPLC.
   RI-700NEXT
Measurement Method Deflection
Flow Cell 8µl volume
RI Range 1.00-1.75
Attenuator 1-2000、∞ in 12 steps



Specifications



LC-9110Ⅱ NEXT LC-9160Ⅱ NEXT
Pump Model P-9101B P-9104B
Pump Type Reciprocating double plunger
 Flow Rate 0.01 - 10.00ml /min 0.01 - 60.00ml /min
Maximum Discharge Pressure 30MPa(0.01 - 5.00ml /min)
25MPa(5.01 - 10.00ml /min) 
20MPa(0.01 - 30.00ml /min)
10MPa(30.01 - 60.00ml/min)
Plunger Φ4.0mm Φ4.0mm
Stroke 8mm
Discharge rate per stroke 100µl 300µl
Flow Stability CV less than 0.3
Operation Panel 8.4inch LCD touch-sensitive screen
ISO/GRA Mode Switching Program and display on LCD touch-sensitive screen
Gradient Method Low-pressure mixture method
Gradient solvent 4 solvents
Pressure Display & Limiter Program and display on LCD touch-sensitive screen
Variable Injector (Manual) Rheodyne 7725i
Sample Loop 3ml (standard)
5ml, 10ml (optional) 
10ml (standard)
5ml, 15ml (optional) 
Event Marker Injection, Recycle and Collect
Recycle System * With non-Diffusion function Recycling system (patent) 
Recycle Method * Swiching Scheme of Solenoid Valve
Auto Recycler * Built-in (1 to 99 times repetition possible) 
Repeat Injector * Direct auto sample loading system (100µl to 99.9ml)
Manual Recycle  Program and display on LCD
Fraction Collector  Standard accessary. Max 33 fraction (Manual /Time mode)
Auto Rerun Function  Save previous operation and rerun up to 99 times.
Injection Overlap Function Time saving auto injector
Auto Cleanup Function  Flow Channel washing by program
Dimensions of Mainframe (mm) 464(W)x492x(H)x504(D)
Tray dimension for fraction collector 440(W)x325(D)
Solvent rack demension for Φ40mm column(s) 240(W)x650x(H)x470(D)
Power Supply  85V - 240V Power consumption 800VA 
Weight  Approx. 40kg